Half Bridge Low Profile Medical Micro Load Cells Side Mounted CZL928F For Body Scale
Product Details:
Place of Origin: | DONGGUAN,GUANGDONG,CHINA |
Brand Name: | UCHI |
Certification: | CE, ISO9001: 2008 |
Model Number: | CZL928F |
Payment & Shipping Terms:
Minimum Order Quantity: | 500pcs |
---|---|
Price: | Negotiable |
Packaging Details: | Bulk |
Delivery Time: | 14 workdays |
Payment Terms: | T/T, Western Union, MoneyGram |
Supply Ability: | 500000pcs |
Detail Information |
|||
Product Name: | Medical Load Cells | Rated Load: | 50KG |
---|---|---|---|
Protection: | IP65 | Application: | Body Scales |
PN: | CZL928F | HS Code: | 9031809090 |
High Light: | load cell miniature,small load cell |
Product Description
Half Bridge Low Profile Medical Micro Load Cells Side Mounted CZL928F For Body Scale
Introduction
A load cell is a force sensing module - a carefully designed metal structure, with small elements called strain gauges mounted in precise locations on the structure. Load cells are designed to measure a specific force, and ignore other forces being applied. The electrical signal output by the load cell is very small and requires specialized amplification. Fortunately, the 1046 PhidgetBridge will perform all the amplification and measurement of the electrical output.
Load cells are designed to measure force in one direction. They will often measure force in other directions, but the sensor sensitivity will be different, since parts of the load cell operating under compression are now in tension, and vice versa.
This Single Point Load Cell is used in small jewelry scales and kitchen scales. It’s mounted by bolting down the end of the load cell where the wires are attached, and applying force on the other end in the direction of the arrow. Where the force is applied is not critical, as this load cell measures a shearing effect on the beam, not the bending of the beam. If you mount a small platform on the load cell, as would be done in a small scale, this load cell provides accurate readings regardless of the position of the load on the platform.
Product specifications
1. Rated Load | 50 (Kg) |
2. Rated Output | 1.0±0.15 mV/V |
3. Zero Balance | +/-0.5mV/V |
4. Input Resistance | 1000±10Ω |
5. Output Resistance | 1000±10Ω |
6. Excitation Voltage | 3-10VDC |
7. Nonlinearity | 0.3%F.S |
8. Hysteresis | 0.3%F.S |
9. Repeatability | 0.2%F.S |
10.Creep(30min) | 0.1%F.S |
11.Operating Temperature | -10℃ To +60℃ |
12.Temperature Effect on Zero | 0.3%F.S/10℃ |
13.Temperature Effect on Span | 0.3%F.S/10℃ |
14.Insulation Resistance | 2000MΩ(50VDC) |
15. Four Angle Error | +/-0.1%F.S |
16.Ultimate Overload | 150%FS |
NumberType="1" TCSC="0" w:st="on">5mm 4-wire shielding cable)
Installation
A single-point load cell. Applying force to one end creates stress in the bar, causing the strain gauge encased in the white plastic to flex.This Single Point Load Cell is used in small jewelry scales and kitchen scales. It’s mounted by bolting down the end of the load cell where the wires are attached, and applying force on the other end in the direction of the arrow. Where the force is applied is not critical, as this load cell measures a shearing effect on the beam, not the bending of the beam. If you mount a small platform on the load cell, as would be done in a small scale, this load cell provides accurate readings regardless of the position of the load on the platform.
Button Load Cells
A button load cell. Applying force to the bump in the middle of the cell causes a change in the output.This Button Load Cell is used in applications that require a thin form factor. The bottom of the load cell is bolted, and force applied to the button on the top. By loading only the button, which is slightly rounded, the load cell is less sensitive to errors resulting from the load not pushing down exactly straight on the load cell.